Use este identificador para citar ou linkar para este item: http://repositorio.bc.ufg.br/handle/ri/17606
Tipo do documento: Artigo
Título: Axonal electrovisiogram as an electrophysiological test to evaluate optic nerve and inner retina electrical potentials: findings in normal subjects
Título(s) alternativo(s): Eletrovisuograma axonal como teste eletrofisiológico na avaliação dos potenciais de ação do nervo óptico e da retina interna: achados em indivíduos normais
Autor: Cella, Wener Passarinho
Dantas, Adalmir Mortera
Lima, Alexandre Vasconcelos
Ávila, Marcos Pereira de
Resumo: Objetivos: Padronizar e validar a técnica de realização do eletrovisuograma axonal (EVA), definir seus valores normativos e caracterizar os achados em indivíduos normais. Métodos: Estudo descritivo com 140 indivíduos (280 olhos) normais. Os partici- pantes foram divididos em sete grupos de acordo com a idade, cada qual com 10 indivíduos do sexo masculino e 10 do sexo feminino. Definiu-se como técnica de exame a estimulação monocular por flash luminoso com intensidade de 0 dB na cúpula de Ganzfeld a uma frequência de 1,4 Hz. Foram utilizados eletrodos com cúpula de ouro e foram analisados os traçados elétricos obtidos após rejeição de artefatos. Para cada amplitude e tempo de culminação foram calculados a média, a mediana, o desvio-padrão, os valores mínimo e máximo e o intervalo de confiança de 95% representando uma faixa de normalidade para os valores. Resultados: A técnica de exame padronizada foi a estimulação visual monocular por flash em condições mesópicas. O traçado normal do eletrovisuograma axonal consis- tiu numa onda positiva inicial (P1, com amplitude média de 2,0 mV e tempo de culminação médio de 23,1 ms) seguida de uma onda negativa (N1, com amplitude média de -3,9 mV e tempo de culminação médio de 41,4 ms). Não foram observadas diferenças significativas entre os sexos e entre os olhos direito e esquerdo, mas os tempos de culminação de P1 e de N1 aumentaram proporcionalmente com a idade. Conclusões: O eletrovisuograma axonal é um exame de potencial visual evocado pré-quiasmático, capaz de registrar de forma confiável e reprodutível a atividade elétrica do nervo óptico e da retina interna, podendo ser utilizado na propedêutica eletrofisiológica na investigação de lesões neurorretinianas.
Abstract: Purposes: To standardize and validate the technique of axonal electrovisiogram (AxEvg), defining its normative values and parameters and characterizing its findings in normal individuals. Methods: We enrolled 140 normal individuals (280 eyes) divided into seven groups according to age, each one with 10 males and 10 females. The technique was based on monocular visual stimulation by a 0 dB intensity bright flash on Ganzfeld bowl at a presentation rate of 1.4 Hz. Golden cup electrodes were used and electrical waves were acquired after artifact rejection. For each amplitude and implicit time peak we calculated the mean, median, pattern deviation, minimum and maximum values and 95% confidence interval. Results: Monocular visual stimulation with bright flash under mesopic conditions was the standard technical procedure established. The normal AxEvg waveform consists of an initial positive wave (named P1, with mean amplitude of 2.0 mV and mean implicit time peak of 23.1 ms) followed by a negative wave (named N1, with mean amplitude of -3.9 mV and mean implicit time peak of 41.4 ms). No significant differences were observed between males and females or between right and left eyes, but there was an increased P1 and N1 implicit time peaks according to age. Implicit time characteristics suggest that P1 wave represents an optic nerve electrical potential and N1 wave represents an inner retinal layers potential. Conclusions: AxEvg can be considered a pre-chiasmatic visual evoked potential capable to reliably record the electrical activity of optic nerve and inner retina. The findings suggest that AxEvg may be useful as an electrophysiological test in the diagnosis of neuroretinal diseases.
Palavras-chave: Electrophysiology
Retina
Electroretinography
Electrodiagnosis
Evoked potentials, visual
Photic stimulation
Optic nerve/physiology
Eletrofisiologia
Eletrorretinografia
Eletrodiagnóstico
Retina
Potenciais evocados visuais
Nervo óptico/fisiologia
Estimulação luminosa
País: Brasil
Unidade acadêmica: Faculdade de Medicina - FM (RG)
Citação: CELLA, Wener Passarinho et al. Axonal electrovisiogram as an electrophysiological test to evaluate optic nerve and inner retina electrical potentials: findings in normal subjects. Arquivos Brasileiros de Oftalmologia, Rio de Janeiro, v. 74, n. 1, p. 37-43, 2011.
Tipo de acesso: Acesso Aberto
Identificador do documento: 10.1590/S0004-27492011000100009
Identificador do documento: 10.1590/S0004-27492011000100009
URI: http://repositorio.bc.ufg.br/handle/ri/17606
Data de publicação: 2011
Aparece nas coleções:FM - Artigos publicados em periódicos

Arquivos associados a este item:
Arquivo Descrição TamanhoFormato 
Artigo - Wener Passarinho Cella - 2011.pdf426,39 kBAdobe PDFThumbnail
Baixar/Abrir


Este item está licenciada sob uma Licença Creative Commons Creative Commons