Visible LED-based instrumentation for photometric determination of electroosmotic flow in microchannels
| dc.creator | Silva, José Alberto Fracassi da | |
| dc.creator | Deblire, Ariane | |
| dc.creator | Jesus, Dosil Pereira de | |
| dc.creator | Coltro, Wendell Karlos Tomazelli | |
| dc.date.accessioned | 2019-01-23T14:44:11Z | |
| dc.date.available | 2019-01-23T14:44:11Z | |
| dc.date.issued | 2011-01 | |
| dc.description.abstract | An instrumental setup for electroosmotic flow (EOF) determination in microfluidic devices is described. The system is based on a trinocular microscope and an integrated photodiode OPT101. The radiation was provided by a high-intensity white LED. For the determinations, patent blue V and methylene blue aqueous solutions were used. The setup was applied for EOF monitoring in hybrid toner/glass and PDMS microchips. For the toner/glass device, the EOF mobility was determined to be (5.75 ± 0.01) 10-4 cm2 V-1 s-1 at pH 7.0. For PDMS devices the measured EOF mobility was (3.2 ± 0.1) 10‑4 cm2 V‑1 s‑1. Reproducible measurements were obtained in all experiments, which produced results with small errors. The proposed method was compared to the conventional current monitoring method. | pt_BR |
| dc.description.resumo | A instrumentação para a determinação fotométrica do fluxo eletrosmótico (EOF) em dispositivos microfluídicos é descrita neste trabalho. A instrumentação é baseada em componentes acessíveis e consiste em um microscópio trinocular e no fotodiodo integrado OPT101. Um diodo emissor de luz (LED) de alta intensidade foi utilizado como fonte de radiação. Para as determinações foram utilizadas soluções aquosas dos corantes azul patente V e azul de metileno. O sistema foi utilizado no monitoramento do EOF em microdispositivos de poli(dimetilsiloxano) (PDMS) e híbridos de toner/vidro. As mobilidades do EOF determinadas em pH 7,0 foram (5,75 ± 0,01)10-4 cm2 V-1 s-1 e (3,2 ± 0,1)10‑4 cm2 V‑1 s‑1 para os dispositivos de toner/vidro e PDMS, respectivamente. Medidas reprodutíveis foram obtidas em todos os experimentos, levando a uma alta precisão na determinação. O método proposto foi comparado com o método tradicional de determinação do EOF que envolve a medida da corrente nos microcanais. | pt_BR |
| dc.identifier.citation | SILVA, José Alberto Fracassi da et al. Visible LED-based instrumentation for photometric determination of electroosmotic flow in microchannels. Journal of the Brazilian Chemical Society, Campinas, v. 22, n. 4, p. 736-740, Jan. 2011. | pt_BR |
| dc.identifier.doi | 10.1590/S0103-50532011000400017 | |
| dc.identifier.issn | 0103-5053 | |
| dc.identifier.issn | e- 1678-4790 | |
| dc.identifier.uri | http://repositorio.bc.ufg.br/handle/ri/16816 | |
| dc.language.iso | eng | pt_BR |
| dc.publisher.country | Brasil | pt_BR |
| dc.publisher.department | Instituto de Química - IQ (RG) | pt_BR |
| dc.rights | Acesso Aberto | pt_BR |
| dc.subject | Electroosmotic flow | pt_BR |
| dc.subject | µTAS | pt_BR |
| dc.subject | Lab-on-a-chip | pt_BR |
| dc.subject | Photometric methods | pt_BR |
| dc.subject | Microfluidic devices | pt_BR |
| dc.title | Visible LED-based instrumentation for photometric determination of electroosmotic flow in microchannels | pt_BR |
| dc.type | Artigo | pt_BR |