Differential pulse voltammetric determination of piroxicam on lanthanide ferric oxide nanoparticles-carbon paste modified electrode

Resumo

Introduction: Piroxicam (PRX) is a widely consumed anti-inflammatory drug, and the assays of its tablets and gels for quality control are mostly performed through High Performance Liquid Chromatography with Ultraviolet Detector (HPLC-UV). Methods: In this regard, an alternate approach, LaFeO3 nanoparticles-modified carbon paste electrode was prepared in order to carry out PRX analysis by using voltammetric approaches. Results: The modified electrode showed better analytical signal. The calibration curve is in the concentration range of 9.00 and 180.00 µmol·L-1 and the limit of detection was obtained as 1.00 µmol·L1 for PRX. The calculated recovery rate was fairly good (99.14%) in comparison with the declared value in the commercial product. Conclusion: As shown by the low overall process cost, performance quickness and analytical results, the proposed method of electroanalysis with the LaFeO3 nanoparticles-modified carbon paste electrode presents a viable and interesting alternative approach to identification and determination of PRX compound in commercial products.

Descrição

Citação

MACEDO, Isaac Yves Lopes de et al. Differential pulse voltammetric determination of piroxicam on lanthanide ferric oxide nanoparticles-carbon paste modified electrode. Current Pharmaceutical Analysis, [s. l.], v. 13, n. 3, p. 271-276, 2018. DOI: 10.2174/1573412913666170410131223. Disponível em: https://www.eurekaselect.com/article/82735. Acesso em: 23 jun. 2023.