Lab-made solid phase microextraction phases for off line extraction and direct mass spectrometry analysis: evaluating the extraction parameters

Nenhuma Miniatura disponível

Data

2019

Título da Revista

ISSN da Revista

Título de Volume

Editor

Resumo

The analyses of drugs and metabolites in complex matrices have been widely studied in recent years. However, due to high levels endogenous compounds and matrix complexity, these analyses require a sample pre-treatment step. To this aim, two lab-made extractive phases were integrated to probe electrospray ionization mass spectrometry (PESI-MS) technique for direct analysis of illicit drugs in biological fluids and phorbol esters in Jatropha curcas extract. The polypyrrole (PPy) phase was electropolymerized onto a platinum wire surface by cyclic voltammetry. The molecularly imprinted polymer (MIP) was synthesized and adhered onto a stainless-steel needle with epoxy resin. The PPy-PESI-MS method showed to be linear in a concentration range from 1 to 500 μg L−¹, with accuracy values between -2.1 and 14%, and precision values between 0.8 and 10.8%. The MIP-PESI-MS method showed to be linear in a concentration range from 0.9 to 30 mg L−¹, with accuracy values between -1.6 and -15.3%, and precision values between 4.1 and 13.5%.

Descrição

Palavras-chave

Citação

BERNARDO, Ricardo Alves et al. Lab-made solid phase microextraction phases for off line extraction and direct mass spectrometry analysis: evaluating the extraction parameters. Journal of Chromatography A, Amsterdam, v. 1603, p. 23-32, 2019. DOI: 10.1016/j.chroma.2019.06.038. Disponível em: https://www.sciencedirect.com/science/article/abs/pii/S0021967319306478?via%3Dihub. Acesso em: 28 jun. 2023.