This report describes the use of toner-based microzone plates for quantitative determination of glucose in artificial urine and human serum samples using colorimetric detection. The proposed approach has exhibited a linear response for glucose concentration levels from 0 to 10 mmol L-1 with determination coefficient (R2) of 0.996. The achieved values for sensitivity and limit of detection were 5.1 AU (mmol L-1)-1 and 0.6 mmol L-1, respectively. The glucose analysis in artificial serum samples revealed error lower than 5% in comparison with certified values provided by the supplier. Lastly, the proposed approach has also provided suitable accuracy (92-105%) for glucose measurements in artificial urine samples.
Descrição
Palavras-chave
Bioanalytic, Clinical assay, Disposable device, Point-of-care testing, Portable instrumentation
Citação
SILVA, Paula B. M.; OLIVEIRA, Karoliny A.; COLTRO, Wendell K. T. Colorimetric Detection of Glucose in Biological Fluids Using Toner-Based Microzone Plates. Journal of the Brazilian Chemical Society, Campinas, v. 28, n. 1, p. 197-201, 2017. DOI: 10.5935/0103-5053.20160164. Disponível em: https://www.scielo.br/j/jbchs/a/rZ5wMNby7tDGmtfgqzRDsTv/abstract/?lang=en. Acesso em: 8 ago. 2024.